Autoclave 75

Autoclave 75


Technical data

Chamber or case: of ss.316-L metrial 3mm thick, 403mm diameter and 700mm height.
Gate: of s.s.316 material, splid and seamless, 8mm thick and 490mm diameter.
Tube and connections of phosphorous, bronzer 316 Or typical steel, anti rust and anti acid.
External body: of sheet with stanic paint
Digital display For temperature, Time and errors.
Temperature can be set until 121.5°c.
Sterilizing time can be set until 99 minutes.
Authorized work pressure (1.2+0.1).
Input voltage: 220V.
Primary current: 15 A.
Consumption power: 2.5 kw.
Frequency: 50 Hz.
Electrically speaking, autoclave is of type 1 and earthed.

Accomplished tests


Hydro sta test: 3bar
Regulate and test the safety valve: 1.6bar
Operation test: 1.3bar
De-electric resistance test: 1650v
Electric eurrent leaking test: 0.05 ma
Pressurized microbial test: 121.5 t 0.1°C